Caire Eclipse 5 Error Codes
If you are experiencing a yellow light, red light, or beeping noise, the following error codes will help determine the cause of the issue. To view the error code on your Eclipse 5, press the no smoking icon on the front panel.
Your machine is working properly.
Cannot charge battery
Warm battery
Ambient pressure fail
You will receive this error when first starting your Eclipse, until the warmup period is completed and the device is producing oxygen greater that 85%. If error code 008 continues after 10 minutes of use, then it is not producing enough oxygen and needs service.
Low battery, 5% or less remaining.
Flow rate error
Loss of power
Malfunction
No breath detected for more than 45 seconds.
Malfunction
Other Alarm Conditions
Possible Cause: The Eclipse 5 is unable to detect the user’s breathing effort.
Your Action:
- Pulse mode may not work correctly if the user has any condition that causes blocked/restricted nasal passages.
- Ensure cannula is no longer than 7 feet in length.
- Ensure that there are no kinks or blockages in cannula tubing.
- If a humidifier bottle is connected, remove the humidifier bottle and connect cannula directly to the outlet port. Pulse mode will not work with a humidifier bottle installed.
Possible Cause: The Eclipse 5 is not receiving power from the attached power supply.
Your Action:
- Verify the outlet is providing power. Do not connect the power supply to a dimmer circuit or a power strip.
- Check that cable connections on power supplies are secure to the wall/vehicle and concentrator. If using the AC power supply, check the cable connection on the power supply transformer, as well.
- Ensure that the power verification lights are on. There will be a green light on the power supply box and there will be a green light on the Eclipse 5 concentrator control panel in the shape of a power cord.
- If power verifications are not lit, remove all connections of the power supply for 20 seconds and reconnect.
Possible Cause #1: The power cartridge was not fully charged
Your Action:
- Connect to AC power to recharge the battery. Verify that the battery charges for 2-5 hours and the battery icon is full and not flashing before use.
Possible Cause #2: The eclipse 5 was not operating at the anticipated flow rate.
Your Action:
- Ensure that you are using the flow rate prescribed by your physician and that pulse or continuous flow is being used as required.
Possible Cause #3: The power cartridge is not performing to specifications.
Your Action:
- Attempt to drain the battery cartridge completely. Do this by running the Eclipse 5 on battery power until the machine shuts off completely. Then remove the battery and allow it to cool for approximately 30 minutes. After this time, re-insert the battery in the Eclipse 5 and recharge it using AC power.
- If the battery still does not last as long as it should after following step 1, then you may need a new battery.
Possible Cause #1: The Eclipse 5 is not receiving external AC or DC power to charge the battery.
Your Action:
- Verify that the AC or DC outlet is providing power. Do not connect the power supply to an AC dimmer or a power strip.
- Check that the cable connections on power supplies are secure to the wall/vehicle and concentrator. If using the AC power supply, check the cable connection on the power supply transformer, as well.
- Ensure that the power verification lights are on. There will be a green light on the power supply box and there will be a green light on the Eclipse 5 control panel in the shape of a power cord.
- If power verification LED’s are not lit, remove all connections of the power supply for 20 seconds and reconnect.
Possible Cause #2: The Eclipse 5 is not properly communicating the power cartridge to charge it.
Your Action:
- Remove power cartridge and re-install to ensure that it is secure in the concentrator.
- Inspect the Eclipse 5 for a solid yellow light when the battery is installed, or a flashing battery icon when attempting to charge. If either of these symptoms is present, continue to step 3.
- If possible, install a separate power cartridge in the Eclipse 5/ If the replacement cartridge works properly, then the original cartridge needs to be replaced.